Six, correction:<br>Correction is an important measure to ensure the accuracy of the instrument test, the instrument must be in accordance with the following three correction indicators, according to the Pharmacopeia every six months.<br> 1, the accuracy of the sampling volume<br>(1, equipment)<br> Balance (0.01 g accuracy); distilled water; 150ml beaker<br>(2), operation<br>1, power on the instrument.<br>2. Pour distilled water into the beaker (about 2/3) on the inspection table, at this time the agitator should be placed outside the sampling cup.<br>3, point full channel key, set sampling 5 ml, such as the measured volume average of 5 ml to 2% qualified, if you need to adjust the need to use a computer calibration<br>2, the accuracy of particle count (calibration)<br>Example: Correction of the channel of s10 m<br>1, the standard particles: the peak particle size should be within 10 m to 0.5 m;<br>2, the two main peak path should account for more than 90% of the total number of particles;<br>3, dilution concentration: 1000-1500 grains /ml (0.9% sodium chloride injection or clean water).<br>4, correction method: When the instrument is plugged in to complete the self-test, the instrument has been purified water environmental test in accordance with the requirements, the standard particle dilution of 10 m will be turned back and forth 20 times, so that the particle distribution is uniform, take