Time after measurement At the end of the measurement of a running time, the liquid meniscus of the sample will trigger the bottom light sensor. After that, a certain amount of time will pass until the measuring sphere and the capillary tube have been emptied completely; only after that, the sample can be suctioned back into the measuring sphere for a repeated measurement. This waiting period can be changed via this parameter. The value 1 is preset.Sucking time For AVS 370 systems, which are connected to a waste system, the parameter "sucking time"defines the suction time for "Rinsing with next sample". This suction time is the running time of a separate vacuum pump, which suctions the rinsing agent (sample or solvent) into a waste bottle. The suction time is preset to a value of 100 %. In many cases, a shorter time is sufficient for the suction time in order to completely suction off the sample. In these cases. we recommend changing the value to shorten the cleaning process.blowing out before measuring"Blowing out before measuring"can only be activated during the suctioning operating mode. After the activation, the capillary is blown out for about 1 second before the start of the running time measurement. This function is used to prevent air bubbles. Air bubbles can be created when sample residue remains in the capillary after a measurement, which will form air bubbles when the sample is uctioned up with air afterwards. By blowing out the capillaries, this sample residue is removed so that no air bubbles are formed.Enabled sensors There are two safety sensors ("Enabled sensors") that can be connected to the AVS 370: "Safety device"and "Wastebottle". If these sensors are to be used, they must be connected to the AVS 370and activated under "Enabled sensors"The status of the sensors can be checked by clicking the "?"to the right of "Waste bottle". If a safety sensor is connected and has not triggered, it will have a green background, otherwise, it will be red. In the case shown in Fig. 7, the sensor "Safety device"is connected and activated, but not "Wastebottle