The root cause of the deviation is that the calibration staff Chen Hongwei negligently failed to insert the thermometer into the matching temperature sleeve, and H0-TM-21224 "Automatic Control System Temperature Calibration Regulations" did not specify in detail which thermometers correspond to which types of temperature sleeves. <br><br>The mismatch between the thermometer and the temperature sleeve causes the actual error to exceed the allowable error. When recalibrated, the temperature sleeve is replaced with a matching temperature sleeve. The calibration result is qualified. At the same time, the thermometer is removed and placed in the dry body furnace when the 201 can is empty The heating is carried out in the middle, so it does not involve the actual temperature measurement of the tank body, and it will not affect the completed production batch. <br><br>CP-TZ-API-D-TZ-API-EG-19005-20001. Replace the thermometer of Y 201 fermentation 201 tank in workshop 109 with matching temperature sleeve for annual calibration. <br>CP-TZ-API-D-TZ-API-EG-19005-20002. Revision H0-TM-21224 "Automatic Control System Temperature Calibration Regulations" version 02, add "the temperature needs to be put into the thermometer calibration process in the thermometer calibration process" Sleeve, if there is no matching temperature sleeve, it is necessary to add silicone oil, thermal grease and other thermally conductive substances to the temperature sleeve of the dry body furnace ", detailed list of the thermometer and temperature sleeve model comparison table, and training relevant personnel .