[reference interval] 160 normal healthy people were detected, and the 95% percentile of the measured value was: < 120 pg/ml; Each laboratory should determine the applicability of the normal reference interval through experiments, and establish the normal reference interval of this laboratory when necessary. [Explanation of inspection results] Due to the reasons of methodology or antigen specificity, there may be deviations between the detection results of reagents from different manufacturers, so they should not be directly compared with each other to avoid wrong medical explanation. The test results are closely related to the test methods, so the test methods should be indicated in the laboratory results report. Results The diagnosis should be made by combining the patient's medical history and other clinical examination results. When the measured result exceeds the linear range, the sample dilution should be used for measurement, with the maximum dilution factor of 10 times, and the result is the measured value multiplied by the dilution factor; This kit is only used for the determination of serum samples, but the reliability of other body fluid samples has not been fully confirmed by experiments. When the NT-proBNP content of the sample reaches the concentration level of 200000pg/mL, the measured value is still greater than the upper linear limit (30,000 pg/ml), which has no influence on the actual detection judgment.