The working LOQ of the method should be evaluated in an analytical run. <br>The LOQ should be determined by preparing three samples in parallel from a standard solution, the SN of the peak area obtained from this sample solution should be about 10.0, and each sample prepared should be injected once. <br>The peak response should be calculated based on the calibration curve, and the measured average calculated concentration must be within the range of 100.0±25.0% of the labeled concentration.