Keep the parameter settings of the DSA machine fixed, and set the vertical Y-axis values of the detector coordinates to 0, 30, 60, 90, 120, 150, and 180 on the left and right vertical detection planes of the bed unit respectively. Corresponding to each vertical Y-axis value, adjust the horizontal X-axis value and set it to 0, 30, 60, 90, 120, 150, 180, 210, 240, 270, 300 in turn. The coordinate 0 point is located at the intersection of the head end of the bed unit and the ground level, and the radiation detector is placed at the intersection of each XY axis value, so that the radiation detector ray acquisition window faces the bed unit. 77 radiation parameters can be obtained corresponding to each vertical detection plane.