To start, on the 'V.A.C.VeraFlo™ Settings' screen select 6mL for the 'Instill Volume' and select 1s for the 'Soak Time'.All other settings can be left as default.Note:The V.A.C. time needs to be greater than 60 seconds.VeraFlo Settings Screen, Select Instill Volume and DurationSelect '01c' at the bottom of the screen.VeraFlo™ Settings Screen, Select SettingsSince this is the first time therapy is activated, the therapy unit will need to establish the dressing seal.Once the Seal Check™ Leak Detector has finished running, the therapy unit will instill fluid.Note:If additional fluid has entered the canister since step 4.1 or if the scale has appeared to drift, repeat step 3c on the Flow Rate and Accuracy instructions prior to beginning the instillation cycle.Seal Check™ Leak Detector ScreenExecute step 4c of the Flow Rate and Accuracy instructions and stop the data capture after the fluid is instilled into the fluid collection container.Record the volume and duration from the Labview program on Table 4 of the Test Results Form following these instructions.Additionally, enter the volume and duration on Table 4 of the electronic copy on the Test Results Form loaded on the laptop.Once the instillation is complete, the therapy unit will switch to V.A.C. therapy.Do not completely stop therapy to c