Next I'll cover the use of pH meters. First of all, I would like to introduce the pH meter, pH meter operator panel buttons are: pH/mV mode toggle button, temperature, slope, positioning adjustment button and confirmation button, and a composite electrode. Preparation: Next, start using the pH meter, remove the electrode protection case, pH composite electrode connected to the pH meter, connect the pH meter power cord, the pH meter preheat more than five minutes. Press the pH/mV button, switch to pH mode, at this time the upper right corner of the instrument screen shows pH, press the temperature button, then press the confirmation, press the upper and lower keys to adjust the temperature to the laboratory ambient temperature, confirm. Remove the appropriate amount of pH meter calibration fluid in the beech, rinse the electrode with distilled water, and dry the electrode with filter paper. Rinse the electrode with pH 6.86 calibration fluid, absorb the liquid on the electrode, open the rubber plug at the rear of the electrode, place the electrode in the pH 6.86 calibration fluid, position it, gently shake the beo cup, press the positioning button and press the confirmation, press the upper and lower keys to adjust the pH to 6.86, and finally press the confirmation button to position it. Rinse the electrode with distilled water and dry it with absorbent paper, rinse the electrode with pH 4.00 calibration fluid, absorb the liquid on the electrode, place the electrode in the pH 4.00 calibration fluid, adjust the slope, shake gently, press the upper and lower keys to adjust pH to 4.00, and finally confirm the button. The calibration is complete and can be used to measure the pH of the acidic solution. To determine the pH of the solution, first rinse the electrode with distilled water, dry it with filter paper, clean it once with the measured liquid, absorb the liquid on the electrode, stir the solution with a glass rod, make the solution even, dip the electrode into the solution under test, shake the beo cup lightly, read it after the number stabilizes, read its pH, measure it three times in parallel, record the data. After testing, clean the electrode, vacuum dry with filter paper, set up a composite electrode protection case, the set should put a small amount of rehydration liquid (electrode protection fluid), turn off the power supply, unplug the electrode wire, unplug the power cord, set up the instrument, clean up the table. The introduction to the use of the pH meter is complete, thank you for listening.
正在翻译中..