Locking detection of locking lever<br> Install the evaporator to be tested on PENLON evaporator test fixture (0050-J19) and hang it.<br> The locking lever can be locked smoothly by rotating the interlocking cover clockwise by about 90; Turn the interlocking cover counterclockwise by about 90 to open it smoothly;<br> If this is repeated for 3 times, it can be locked and opened smoothly, then the locking function of the surface locking rod meets the requirements, otherwise it is unqualified.<br> Zero button detection<br> Do not press the zero button, and check that the evaporator handwheel cannot be opened when it rotates counterclockwise.<br> Press the zero button and check that the evaporator handwheel can be opened by rotating counterclockwise.<br> Liquid level display tube inspection<br> For all configured evaporators, it is necessary to check whether the liquid level display tube is installed correctly to ensure that it is not installed backwards; The correct installation is shown in Figure 3A and B (take Key Filler as an example).<br> 1 scale<br> From the front and left side, you can see a vertical line in the middle of the liquid level pipe.<br> Liquid level tube of dosing device<br> <br> Figure iii a<br> Figure iii b<br> Air tightness detection at the maximum scale<br> Connect the gas path comprehensive test fixture (high pressure) (0611-TF10) to the inlet of the flowmeter (suitable for high pressure), and connect the outlet of the flowmeter to the PENLON evaporator test fixture (0050-J19), and plug the air outlet of the volatilization tank support assembly on the fixture with a pipe plug.<br> (Choose one method according to available tooling).<br> Install the evaporator to be tested on PENLON evaporator test fixture (0050-J19) and lock the locking lever.<br> Confirm that the tooling power supply, tooling air supply, main switch and exhaust switch have been turned on; Turn the evaporator handwheel to the maximum scale.<br> Adjust the pressure regulating valve of the tooling to make the displayed value of the pressure gauge (50 1) kPa. After the reading of the pressure gauge is stable, observe that the leakage displayed by the flowmeter (applicable to high pressure) is ≤ 5ml/min; (Select a test method corresponding to the front end).<br> Fill the above test results into the test record table.<br> Note: After the air tightness test is completed, it is not necessary to close the handwheel. Release the locking lever to take the evaporator to be tested off the test fixture to remove the internal pressure of the evaporator tank, and then rotate the evaporator handwheel to zero.
正在翻译中..