1. For 9051-a191001 ~ 9051-a191009 batch of glucosamine sulfate sodium chloride double salt samples, the test report shall be issued based on the retest results of the original solution, and the original sequence of the atlas shall be discarded and retained.<br>2. After ultrasonic cleaning of the inlet check valve and stainless steel filter head of Shimadzu 53 × liquid phase instrument, the instrument returned to normal.<br>3. Upgraded h0-am-10025 high performance liquid chromatography (version 00): there is a jewel ball in the one-way valve to control the flow direction of the liquid in the pipeline, but the jewel ball is easy to stick with high viscosity solvent such as acetonitrile, resulting in the pipeline blockage. After the instrument is tested with 100% pure acetonitrile, the pipeline needs to be flushed with hot water at about 60 ℃. And train the relevant personnel.<br>