After the corresponding detection method is loaded by the software, and the "offset", "arrangement", "background" program is run by the instrument, the analysis begins, the test results are exported, and the D10, D50, and D90 data are collected.<br>3 parallel tests for each weight.<br>The results of RSD measured under the weighing volume of 1.3g, 2g and 3g were 1.4%, 1.2% and 2.2% respectively, and the results all met the standard requirements, and the sample weighing for the test product had no obvious effect on the test results, so the recommended weight of 1.3g was selected as the determination condition.<br>9.1.2 Vacuum<br>Take the test product about 1.3g, precision, set the instrument supporting the cylinder, so that the test product does not excessive barrel bottom arc part, the measuring tube placed in the dry test module (DPS).<br>The D10, D50, D90 data are collected, and D10, D50, and D90 data are collected, respectively, under the conditions of vacuum of 15 " H2O, 20 " H2O, and 30 " H2O (maximum vacuum degree of the instrument).<br>3 in parallel for each vacuum degree.<br>Method parameters:<br>Measuring drift:<br>Measurement background:<br>Refractive index:<br>The results of the rsD of D10, D50 and D90 were 2.0%, 3.1%, and 3.0% respectively under the condition of vacuum of 15 x H2O, 20 x H2O, 30 , H2O, h2O, H2O, H2O, and D90, respectively, indicating that the vacuum degree had no obvious effect on the test results, and the vacuum degree was 15 "H2O to 30" H2O is suitable for microcrystalline cellulose particle size distribution detection.<br>Vacuum degree 30, H2O is the maximum vacuum and default vacuum degree of the instrument, in order to obtain the optimal dispersion effect of the sample, and prevent the vacuum degree from falling below 15 s" H2O, select the maximum vacuum degree of the instrument 30" H2O as the determination condition.
正在翻译中..