T20 ultimately required moderator intervention to turn on the electrosurgical generator (refer to Step 1.14 for associated findings).C9 did not know what “amber button” meant, and the Performance Checklist Workbook did not have an associated image.Insufficient guidance in Performance Checklist Workbook to identify System pause button – The Performance Checklist Workbook uses only written instructions to refer to the “amber button” on the electrosurgical generator, which might cause some users to mistake this button for the system pause button, which is also an amber color (T9, T20).