Measured Points 1Calculated Points 1Integration Time (sec) 8Entrance slit 20Exit Slit 15Increment 0.002Analysis Mode MaxHigh Voltage 990 Gain of PM 100Coating G1 Select OK to return to the previous menu. Select the Parameters tab under this tab and check the following:Rinsing (fixed) 45 sTransfer 20 sStabilization 1 sDelay of synchronization 0sMake changes and Save changes if necessary.3) Go to Sequence Manager.4) Select the Method GRAPABAS in the drop down box located on the top of the screen.5) Select New on the right hand of the screen6) Set up sequence in the following order:a. Peak Searchb. Auto-attenuate c. Calibrationd. Mid/Controle. Unknown sample(s) and blank Note: Periodically through the sequence, setup a mid standard as sample to check instrument drift. Recalibrate if necessary