With the specification at 55 seconds maximum, the 7x7mm deflation time values presented in this report pass with a UTL of 44.9 sec.Since the UTL is below the max specification, it assures with 95% confidence that at least 95% of the population will be within specification.Balloon SealBalloon seal data was not normally distributed due to the small variation in data.Seal test data is recorded as the change in diameter over a 5 minute period.Since the diameter rarely changes at all, most values are zero, thereby preventing the data from being normal.Using Tchebysheff's Theorem with K=5, the calculated UTL for the 4x7mm HyperForm is 0.04mm.