This out of range rule applies only to Available Parameters(see definition NMS‐2519)For the clinical release SW, refer to NMS‐2473 for the cited definition.NGPC‐356 ‐ Detection of Low Lead ImpedanceProgrammer shall notify the user upon detection for the following combination of parameters during interrogation:normal mode output current > 0 mA and measured lead impedance 0 mA and measured lead impedance >=5300 ohms(Refer to High_Lead_Impedance in NGPC‐390 ‐ Notification Text)Measured Lead impedance of >=10000 is displayed as >=10000 Ohms.NGPC‐354 ‐ Interrogation Performance ‐ M102‐M105Once connection is established with IPG, programmer shall interrogate Interrogated Data and display(as per NGPC‐348)within 10 seconds.NGPC‐353 ‐ Interrogation Performance ‐ SR Enabled IPGsOnce connection is established with the IPG, the programmer shall interrogate Interrogated Data and display(as per NGPC‐348)within 30 seconds for at least one interrogation type.Does not apply for M1000 IPGs with day‐night. scheduled, or guided programming mode enabled.NGPC‐348 ‐ Display Interrogated DataProgrammer shall display Therapy Parameters, Detection Parameters, Patient Information and IPG information interrogated from IPG.Interrogation, SecurityOnly Patient ID and implant date are stored and displayed which is not considered a PHI.NGPC‐347 ‐ Interrogate IPGProgrammer shall provide the ability to interrogate Interrogated Data from IPG.NGPC‐51 ‐ IPG compatibilityProgrammer shall support the IPGs as per the Compatibility Matrix.NGPC‐368 ‐ Detection of Non Supported IPGProgrammer shall notify the user upon detection of non Supported IPG.NGPC‐358 ‐ Detection of Empty Patient InformationProgrammer shall notify the user upon detection of empty patient ID and implant date parameters during interrogation."detection of empty patient ID and implant date parameters" can be understood to mean that either field being empty can trigger this condition.NGPC‐811 ‐ Perform Diagnostics Indication ‐ M102/102RFollowing initial interrogation, the programmer shall provide an indication to the user when diagnostics have not been performed for the IPG In‐Session.NGPC‐1622 ‐ Program Autostim FrequencyProgrammer shall program autostim frequency of the IPG In‐Session to be equal to the normal mode frequency in next successful programming event under the following condition:autostim frequency of the IPG is not equal to normal mode frequency.For 106 IPGs:The autostim frequency is to be programmed to 0 so that the IPG actually uses the normal mode frequency for autostim frequency.So to implement this requirement the value for autostim frequency has to programmed to the value 0 for these IPGs.For 1000 IPGs:The autostim frequency value has to be set to normal mode frequency.So to implement this requirement the value for autostim frequency has to programmed to the value of the normal mode frequency for these IPGs.NGPC‐1621 ‐ Program Magnet FrequencyNGPC‐1619 ‐ Program Magnet Off Time‐ 106/1000Programmer shall program magnet off time of the IPG to normal off time in next successful programming event under the following conditions:1)interrogated magnet off time of the IPG In‐Session is not equal to normal off time.2)the user programs a new normal off time into the IPG In‐SessionNGPC‐1618 ‐ Program Magnet Off TimeProgrammer shall program magnet off time of the IPG to 0 in next successful programming event under the following condition:magnet off time of the IPG In‐Session is not equal to 0.103/104/105 IPGs, when magnet off time is set to 0, IPG uses normal off timeNGPC‐1609 ‐ Indication Of The Out Of Range ParametersProgrammer shall indicate the out of range programmable parameters as ?? if the parameter is available to the user.This out of range rule applies only to Available Parameters(see NMS‐2519)NGPC‐1608 ‐ Reset Pulse Disabled FlagProgrammer shall reset the pulse disabled flag in the IPG In‐Session in the next successful programming event.Reset means that the PulseDisabled flag is set to a value of 0.NGPC‐1519 ‐ Programming Of Autostim ParametersProgrammer shall provide ability to program autostim ouput current, autostim on time, autostim pulse width when tachycardia/seizure detection is enabled in the IPG In‐Session AND user has not selected tachycardia/seizure detection to program to be disabled within current session.This applies to both Daytime and Nighttime set when the Day‐Night programming is enabledNGPC‐1388 ‐ Programming Of Detection ParametersProgrammer shall provide ability to program Detection Parameters(except tachycardia/seizure detection)when tachycardia/sizure detection is enabled in the IPG In‐Session AND user has not selected tachycardia/seizure detection to program to be disabled within current session.NG