15:00, an IPC laboratory analyst He Qinqin microorganisms in HPLC (ID: CAY-WSW1-09) sample spectra processing of pravastatin sodium "3036-E1190403 a purified product" on the instrument, found Rt: 11.591 the peak occurs at the flat min. During sampling, He Qinqin found there may be small air bubbles into the injector, but the data has been collected, it can not be stopped. <br><br>The reason for this deviation occurs when the analyst He Qinqin manual injection, the sample "3036-E1190403 a purified product" inject injector, due to the air bubbles injected into the barrel, Rt: 11.591min flat peak at bubbles peak, regardless of the sample itself, has no effect on product quality. <br><br>The only variation relates to a sample of pravastatin sodium "3036-E1190403 a purified product", through investigation that, since the sample was injected manually when the analyst injector, due to the air bubbles injected into the barrel. After finding laboratory error correction re-testing results without exception, and as the result of test report issued. The bias had no effect on the production process and product quality. <br><br>1. void and retain sample "3036-E1190403 a purified product" in the initial test pattern. <br>2. After correction re-testing of liquid samples "3036-E1190403 once refined products" and thus test report issued. <br>3. In this case the deviation of the column, manual analysis of all the training phase H3-TM-22108 "10A / 20A Series HPLC protocol" version of the injector 03 Operating Details "is injected into the sample injection when the sample loop, in particular the need to observe the presence or absence of small bubbles into the syringe, prevent bubble injection. "