This deviation is due to the standard product administrator for a moment's negligence, take the wrong standard product, and the inspector Wang Wei did not check the label information, resulting in the use of standard products error, resulting in peak time anomalies, is a human error. <br><br>1. This deviation is caused by human error, due to the use of standard products lead to the standard peak time anomaly, deviation map and the original liquid retest results of impurity peak consistent, but the experimental results can not be calculated, this experimental map waste reservation for reference only, to sample the original liquid retest results issued a test report, this deviation has no effect on product quality evaluation.<br>2. The experimenter's experiment in the last year did not make a similar error, and it was the first time in the last two years that the laboratory had the wrong deviation in the use of standard products. If the standard product slots incorrectly, it will be found in time, the risk can be controlled, so the product batch quality risk detected by the experimenter in the early stage can be controlled.<br>3. Check the standard product administrator Cheng Wenxian, is also the first time such deviation occurred, the risk can be controlled.<br><br>1.HS-25-I three batches of sample experimental results are invalid, with the original liquid retest results issued a test report.<br>2. Training of the relevant personnel in H0-SOP-72013 Standard Management Procedure 01, emphasizing that the analysts need to check that the packaging of the standard products is in good condition before using the standard products and that the information in the label is correct.
正在翻译中..