Practically, the creep phase of a nanoindentation creep test is limited to 70 seconds partly because, relative to sample displacement rates, displacement piezo signal drift rates become amplified over longer data acquisition durations. While 10 seconds is insufficiently low to characterize any meaningful creep, hold times around 70 seconds still maintain LVE and are within the displacement piezo ranges of the data acquisition system.