Keep the parameter settings of the DSA machine fixed, and set the vertical Y-axis values of the detector coordinates to 0, 30, 60, 90, 120, 150, and 180 in the two vertical detection planes of the head and foot of the bed unit respectively. Corresponding to each vertical Y-axis value, adjust the horizontal X-axis value, and set it to 0, 30, 60, 90 in turn. The coordinate 0 points of the two vertical detection planes are located at the intersection of the left side of the head end of the bed unit with the ground level and the left side of the foot end with the ground level. The radiation detector is placed at the intersection of each XY axis value, so that the radiation The detector ray acquisition window faces the bed unit. 28 radiation parameters can be obtained corresponding to each vertical detection plane.